Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
1.5K
Pedicle screw augmentation with bone cement enforced Vicryl mesh.
Samuel L Schmid1, Elias Bachmann1,2, Michael Fischer1
1Department of Orthopedics, Balgrist University Hospital, University of Zürich, Forchstrasse 340, Zürich 8008, Switzerland.
Summary
Augmenting pedicle screws with bone cement and Vicryl mesh (BCVM) reduced cement extrusion compared to standard bone cement (BC). While BC showed higher pullout strength in saw bone blocks, BCVM offered comparable strength in cadaveric vertebrae with improved safety.
Area of Science:
- Orthopaedic Research
- Biomaterials Engineering
- Spinal Surgery
Background:
- Securing pedicle screws in osteoporotic or previously operated bone presents significant challenges.
- Existing augmentation techniques, like different screw types or cement augmentation, often lead to complications and difficult revisions.
- There is a need for improved methods to enhance pedicle screw fixation while minimizing risks.
Purpose of the Study:
- To evaluate the efficacy of biocompatible textile materials (Vicryl mesh) combined with bone cement for augmenting pedicle screw pullout strength.
- To assess the impact of this technique on reducing cement extrusion.
- To compare the biomechanical performance of standard bone cement versus textile-reinforced bone cement.
Main Methods:
- A biomechanical trial involving pedicle screws augmented with standard bone cement (BC) or bone cement reinforced with Vicryl mesh (BCVM).
- Testing was conducted in osteoporosis-like saw bone blocks and human cadaveric lumbar vertebrae.
- Pullout testing was performed to measure screw fixation strength, and cement extrusion was monitored.
Main Results:
- In saw bone blocks, standard bone cement (BC) demonstrated significantly higher mean pullout force (350 N) than the BCVM technique (240 N).
- In cadaveric vertebrae, mean pullout forces were comparable between BC (784 N) and BCVM (757 N), with no statistical difference.
- Crucially, cement extrusion occurred in 40% of BC cases but was completely avoided with the BCVM technique.
Conclusions:
- Textile reinforcement of bone cement for pedicle screw augmentation effectively reduces cement extrusion compared to conventional methods.
- The BCVM technique provides reproducible mechanical strength, comparable to standard bone cement in cadaveric specimens.
- This approach offers a safer alternative for enhancing pedicle screw fixation in challenging bone conditions.