Assessment of peak skin dose in interventional cardiology: A comparison between Gafchromic film and dosimetric
J Greffier1, C Van Ngoc Ty2, G Bonniaud2
1Department of Radiology, Nîmes University Hospital, Medical Imaging Group Nîmes, EA 2415, Bd Prof Robert Debré, 30029 Nîmes Cedex, France.
Objective:
To compare the use of a dose mapping software to Gafchromic film measurement for a simplified peak skin dose (PSD) estimation in interventional cardiology procedure.
Methods:
The study was conducted on a total of 40 cardiac procedures (20 complex coronary angioplasty of chronic total occlusion (CTO) and 20 coronary angiography and coronary angioplasty (CA-PTCA)) conducted between January 2014 to December 2015. PSD measurement (PSDFilm) was obtained by placing XR-RV3 Gafchromic under the patient's back for each procedure. PSD (PSDem.dose) was computed with the software em.dose©. The calculation was performed on the dose metrics collected from the private dose report of each procedure. Two calculation methods (method A: fluoroscopic kerma equally spread on cine acquisition and B: fluoroscopic kerma is added to one air Kerma cine acquisition that contributes to the PSD) were used to calculate the fluoroscopic dose contribution as fluoroscopic data were not recorded in our interventional room. Statistical analyses were carried out to compare PSDFilm and PSDem.dose.
Results:
The PSDFilm median (1st quartile; 3rd quartile) was 0.251(0.190;0.336)Gy for CA-PTCA and 1.453(0.767;2.011)Gy for CTO. For method-A, the PSDem.dose was 0.248(0.182;0.369)Gy for CA-PTCA and 1.601(0.892;2.178)Gy for CTO, and 0.267(0.223;0.446)Gy and 1.75 (0.912;2.584)Gy for method-B, respectively. For the two methods, the correlation between PSDFilm and PSDem.dose was strong. For all cardiology procedures investigated, the mean deviation between PSDFilm and PSDem.dose was 3.4±21.1% for method-A and 17.3%±23.9% for method-B.
Conclusion:
The dose mapping software is convenient to calculate peak skin dose in interventional cardiology.
More Related Videos
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Biological Effects of Radiation
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
X-ray Imaging