Effects of a closed system suction connector on airway resistance in ventilated neonates
Sezin Ünal1, Ebru Ergenekon1, Ebru Kazanci1
1Division of Neonatology, Department of Pediatrics, Gazi University Hospital, Ankara, Turkey.
Background/Aim:
Increased airway resistance reduces the effectiveness of ventilation treatment. Endotracheal tubes (ETTs) and connectors contribute to resistance. However, the effect of a closed system suction (CSS) connector is not well known. We compared the in vivo resistance occurring with a CSS connector with that of the standard connector.
Materials And Methods:
This prospective study was conducted at Gazi University Hospital's neonatal intensive care unit. Intubated neonates were studied for two cycles; each cycle contained two periods of ETT + connector pairs (15 min/period) as follows: cycle 1 [A: long ETT + standard connector; B: long ETT + CSS connector] and cycle 2 [C: shortened ETT + standard connector; D: shortened ETT + CSS connector]. Resistance of 40 breaths/period was averaged for each case, and the means were analyzed by Wilcoxon test for pairwise comparisons between standard and CSS connectors. As each case provided two cycle data, 16 cycle data were compared.
Results:
The CSS connector increased resistance by 13.8% (range: 3.0%-22.1%) compared to the standard connector; P < 0.001. The resistance increase was similar between long [17.3% (range: 3.0%-17.7%)] and shortened ETTs [15.3% (range: 5.0%-29.6%)]; P = 0.834.
Conclusion:
CSS connectors were found to increase airway resistance in ventilated neonates. The contribution of CSS should be considered during ventilation, particularly in the presence of difficulty in providing sufficient tidal volume.
More Related Videos
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
08:58Development of a Neonatal Piglet Acute Lung Injury Model Recreating the Early Environment of Preterm Infant Lungs
Published on: October 31, 2025
Related Concept Videos
Suctioning the Nasopharyngeal Airway
Equipment Required
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...