Method for lateral force calibration in atomic force microscope using MEMS microforce sensor
Related Experiment Video
Updated: Feb 28, 2026

Covalent Immobilization of Proteins for the Single Molecule Force Spectroscopy
Published on: August 20, 2018
Cezary Dziekoński1, Wojciech Dera1, Dariusz M Jarząbek1
1Polish Academy of Sciences, Institute of Fundamental Technological Research, Warsaw, Poland.
Abstract:
In this paper we present a simple and direct method for the lateral force calibration constant determination. Our procedure does not require any knowledge about material or geometrical parameters of an investigated cantilever. We apply a commercially available microforce sensor with advanced electronics for direct measurement of the friction force applied by the cantilever's tip to a flat surface of the microforce sensor measuring beam. Due to the third law of dynamics, the friction force of the equal value tilts the AFM cantilever. Therefore, torsional (lateral force) signal is compared with the signal from the microforce sensor and the lateral force calibration constant is determined. The method is easy to perform and could be widely used for the lateral force calibration constant determination in many types of atomic force microscopes.
More Related Videos
10:55Extracting the Young's Modulus of Native Murine Pulmonary Basement Membranes from Atomic Force Microscopy Derived Force Maps
Published on: January 31, 2025
11:18Use of Atomic Force Microscopy to Measure Mechanical Properties and Turgor Pressure of Plant Cells and Plant Tissues
Published on: July 15, 2019