Mid-term results of percutaneous ventricular septal defect closure with Amplatzer Duct Occluder-II in children

Ozge Pamukcu1, Nazmi Narin1, Ali Baykan1

  • 1Division of Pediatric Cardiology,Erciyes University School of Medicine,Kayseri,Turkey.

Insights

Percutaneous closure of ventricular septal defects using the Amplatzer Duct Occluder-II device is safe and effective in children. This study shows positive mid-term results, even for infants under one year old.

Area of Science:

  • Pediatric Cardiology
  • Interventional Cardiology
  • Medical Devices

Background:

  • Percutaneous closure of ventricular septal defects (VSDs) is an emerging alternative to surgical repair.
  • The development of an ideal device for percutaneous VSD closure remains an ongoing challenge.

Purpose of the Study:

  • To evaluate the mid-term outcomes of percutaneous VSD closure using the Amplatzer Duct Occluder-II (ADO-II) device in pediatric patients.
  • To assess the safety and efficacy of the ADO-II device, particularly in infants.

Main Methods:

  • A retrospective analysis of 49 pediatric patients who underwent percutaneous VSD closure with the ADO-II device between April 2011 and October 2016.
  • Included patients with muscular and perimembranous VSDs, with seven patients younger than one year.

Main Results:

  • The mean patient age was 86.8 months, and mean weight was 24.3 kg. Mean defect diameter was 3.7 mm.
  • Mean fluoroscopy time was 37 minutes and mean procedure time was 74.1 minutes.
  • No major complications occurred during procedures; one patient developed complete atrioventricular block at 6 months follow-up. Total follow-up was 66 months.

Conclusions:

  • The Amplatzer Duct Occluder-II device demonstrates safety and effectiveness for percutaneous VSD closure in a large pediatric series.
  • The device is suitable for use in infants under one year of age, based on the 66-month follow-up data.