Goal-Directed Perfusion Methodology for Determining Oxygenator Performance during Clinical Cardiopulmonary Bypass.
Alfred H Stammers1, Randi Miller2, Stephen G Francis3
1SpecialtyCare, Nashville, Tennessee.
Newer oxygenators with integrated filters performed comparably to older models. The Terumo SX18 oxygenator showed similar gas exchange and hemodilution effects to newer devices like the LivaNova Inspire6F and Terumo FX15.
Area of Science:
- Cardiovascular Surgery
- Biomedical Engineering
- Medical Devices
Background:
- New oxygenator designs integrate arterial line filtration, potentially impacting gas exchange.
- Evaluating the clinical performance of these integrated filtration systems is crucial.
Purpose of the Study:
- To compare the gas exchange performance and clinical outcomes of three different oxygenator models during cardiopulmonary bypass (CPB).
- To assess the effect of integrated versus external arterial line filtration on patient parameters.
Main Methods:
- A prospective clinical study involving 60 adult cardiac surgery patients.
- Patients were divided into three groups based on oxygenator type: Terumo SX18 (integrated filter), Terumo FX15 (external filter), and LivaNova Inspire6F (integrated filter).
- Standardized goal-directed perfusion and patient management protocols were applied.
Main Results:
- The LivaNova Inspire6F required lower fraction of inspired oxygen compared to both Terumo models.
- The Terumo FX15 showed slightly higher arterial oxygen content and delivery.
- Postoperative transfusion rates were higher in the Terumo SX18 group compared to the other two groups.
Conclusions:
- The Terumo SX18 oxygenator demonstrated comparable gas exchange performance and hemodilution to newer generation oxygenators.
- Integrated arterial line filtration in oxygenators did not lead to significant differences in postoperative morbidity.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
06:22Author Spotlight: Advancing Cardiovascular Research — Tailored Langendorff Perfusion Techniques for Improved Experimental Outcomes
Published on: June 14, 2024
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...