Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Comparison of the temperature and humidity in the anesthetic breathing circuit among different anesthetic
Yoon Ji Choi1, Sam Hong Min, Jeong Jun Park
1Department of Anesthesia and Pain Medicine, Pusan National University Yangsan Hospital, Yangsan, Gyeongsangnam-do TnTn Hospital Department of Anesthesiology and Pain Medicine, College of Medicine, Korea University, Seoul, Korea.
Background:
For patients undergoing general anesthesia, adequate warming and humidification of the inspired gases is very important. The aim of this study was to evaluate the differences in the heat and moisture content of the inspired gases with low-flow anesthesia using 4 different anesthesia machines.
Methods:
The patients were divided into 11 groups according to the anesthesia machine used (Ohmeda, Excel; Avance; Dräger, Cato; and Primus) and the fresh gas flow (FGF) rate (0.5, 1, and 4 L/min). The temperature and absolute humidity of the inspired gas in the inspiratory limbs were measured at 5, 10, 15, 30, 45, 60, 75, 90, 105, and 120 minutes in 9 patients scheduled for total thyroidectomy or cervical spine operation in each group.
Results:
The anesthesia machines of Excel, Avance, Cato, and Primus did not show statistically significant changes in the inspired gas temperatures over time within each group with various FGFs. They, however, showed statistically significant changes in the absolute humidity of the inspired gas over time within each group with low FGF anesthesia (P < .05). The anesthesia machines of Cato and Primus showed statistically significant changes in the absolute humidity of the inspired gas over time within each group with an FGF of 4 L/min (P < .05). However, even with low-flow anesthesia, the temperatures and absolute humidities of the inspired gas for all anesthesia machines were lower than the recommended values.
Conclusion:
There were statistical differences in the provision of humidity among different anesthesia workstations. The Cato and Primus workstations were superior to Excel and Avance. However, even these were unsatisfactory in humans. Therefore, additional devices that provide inspired gases with adequate heat and humidity are needed for those undergoing general anesthetic procedures.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...