Long-term stability of an injection-molded zirconia bone-level implant: A testing protocol considering aging kinetics
Benedikt C Spies1, Madeleine E Maass1, Erik Adolfsson2
1Department of Prosthetic Dentistry, Center for Dental Medicine, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Hugstetter Str. 55, 79106 Freiburg, Germany.
Objective:
Separately addressing the fatigue resistance (ISO 14801, evaluation of final product) and aging behavior (ISO 13356, standardized sample) of oral implants made from yttria-stabilized zirconia proved to be insufficient in verifying their long-term stability, since (1) implant processing is known to significantly influence transformation kinetics and (2) aging, up from a certain level, is liable to decrease fatigue resistance. Therefore, the aim of this investigation was to apply a new testing protocol considering environmental conditions adequately inducing aging during dynamic fatigue.
Methods:
Zirconia implants were dynamically loaded (107 cycles), hydrothermally aged (85°, 60 days) or subjected to both treatments simultaneously. Subsequent, monoclinic intensity ratios (Xm) were obtained by locally resolved X-ray microdiffraction (μ-XRD2). Transformation propagation was monitored at cross-sections by μ-Raman spectroscopy and scanning electron microscopy (SEM). Finally, implants were statically loaded to fracture. Linear regression models (fracture load) and mixed models (Xm) were used for statistical analyses.
Results:
All treatments resulted in increased fracture load (p≤0.005), indicating the formation of transformation induced compressive stresses around surface defects during all treatment modalities. However, only hydrothermal and combinational treatment were found to increase Xm (p<0.001). No change in Xm was observed for solely dynamically loaded samples (p≥0.524). Depending on the variable observed, a monoclinic layer thickness of 1-2μm (SEM) or 6-8μm (Raman spectroscopy) was measured at surfaces exposed to water during treatments.
Significance:
Hydrothermal aging was successfully induced during dynamic fatigue. Therefore, the presented setup might serve as reference protocol for ensuring pre-clinically long-term reliability of zirconia oral implants.