Introduction:
Sigma metrics were applied to evaluate the performance of 20 routine chemistry assays, and individual quality control criteria were established based on the sigma values of different assays.
Methods:
Precisions were expressed as the average coefficient variations (CVs) of long-term two-level chemistry controls. The biases of the 20 assays were obtained from the results of trueness programs organized by National Center for Clinical Laboratories (NCCL, China) in 2016. Four different allowable total error (TEa) targets were chosen from biological variation (minimum, desirable, optimal), Clinical Laboratory Improvements Amendments (CLIA, US), Analytical Quality Specification for Routine Analytes in Clinical Chemistry (WS/T 403-2012, China) and the National Cholesterol Education Program (NECP).
Results:
The sigma values from different TEa targets varied. The TEa targets for ALT, AMY, Ca, CHOL, CK, Crea, GGT, K, LDH, Mg, Na, TG, TP, UA and Urea were chosen from WS/T 403-2012; the targets for ALP, AST and GLU were chosen from CLIA; the target for K was chosen from desirable biological variation; and the targets for HDL and LDL were chosen from the NECP. Individual quality criteria were established based on different sigma values.
Conclusions:
Sigma metrics are an optimal tool to evaluate the performance of different assays. An assay with a high value could use a simple internal quality control rule, while an assay with a low value should be monitored strictly.
Related Concept Videos
Clinically Relevant Drug Product Specifications: Methods of Establishment
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Irritable Bowel Syndrome II: Clinical Features and Diagnostic Evaluation
Irritable Bowel Syndrome (IBS) is classified into subtypes based on the predominant bowel habits as determined by the Bristol Stool Form Scale (BSFS). The subtypes are:
Chemistry of Carbohydrates
Chemistry of Carbohydrates
What is Organic Chemistry?