Background:

  • Intraoperative fluoroscopy, particularly mini-C-arm use, presents potential radiation exposure risks for orthopedic surgeons.
  • Understanding radiation dosimetry for hand and eye is crucial for occupational safety in orthopedic procedures.

Purpose of the Study:

  • To prospectively compare hand versus eye radiation exposure during mini-C-arm fluoroscopy.
  • To evaluate if routine mini-C-arm use results in radiation dosages exceeding established safety limits.

Main Methods:

  • Prospective measurement of hand and eye radiation exposure over 12 months in a single hand surgeon.
  • Data collected from 25 surgical cases utilizing a mini-C-arm fluoroscopy device.

Main Results:

  • Average monthly hand radiation dosage (45.81±14.49 mrem) was significantly higher than eye radiation dosage (<30 mrem).
  • Both hand and eye radiation dosages remained below critical exposure limits set by the International Commission on Radiological Protection.

Conclusions:

  • Routine mini-C-arm utilization in orthopedic surgery does not appear to approach critical radiation exposure levels for the hand or eyes.