Related Experiment Video
Updated: Feb 27, 2026

Filtration Isolation of Nucleic Acids: A Simple and Rapid DNA Extraction Method
Published on: August 6, 2016
Preanalytic process linked to spuriously elevated HIV viral loads: improvement on an FDA-approved process
Gary W Procop1, Alan J Taege2, Colleen Starkey1
1Department of Laboratory Medicine, Cleveland Clinic, Cleveland, OH.
Abstract:
The processing of specimens often occurs in a central processing area within laboratories. We demonstrated that plasma centrifuged in the central laboratory but allowed to remain within the primary tube following centrifugation was associated with spuriously elevated HIV viral loads compared with recentrifugation of the plasma just prior to testing.