Characterisation of the Medipix3 detector for 60 and 80keV electrons
J A Mir1, R Clough1, R MacInnes2
1University of Oxford, Department of Materials, Parks Road, Oxford OX1 3PH, United Kingdom.
Abstract:
In this paper we report quantitative measurements of the imaging performance for the current generation of hybrid pixel detector, Medipix3, used as a direct electron detector. We have measured the modulation transfer function and detective quantum efficiency at beam energies of 60 and 80keV. In single pixel mode, energy threshold values can be chosen to maximize either the modulation transfer function or the detective quantum efficiency, obtaining values near to, or exceeding those for a theoretical detector with square pixels. The Medipix3 charge summing mode delivers simultaneous, high values of both modulation transfer function and detective quantum efficiency. We have also characterized the detector response to single electron events and describe an empirical model that predicts the detector modulation transfer function and detective quantum efficiency based on energy threshold. Exemplifying our findings we demonstrate the Medipix3 imaging performance recording a fully exposed electron diffraction pattern at 24-bit depth together with images in single pixel and charge summing modes. Our findings highlight that for transmission electron microscopy performed at low energies (energies <100keV) thick hybrid pixel detectors provide an advantageous architecture for direct electron imaging.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Determination of Crystal Structures
Mass Analyzers: Common Types
Atomic Emission Spectroscopy: Instrumentation