Image Updating for Brain Shift Compensation During Resection
Related Experiment Video
Updated: Feb 27, 2026

Automated Midline Shift and Intracranial Pressure Estimation based on Brain CT Images
Published on: April 13, 2013
Xiaoyao Fan1, David W Roberts1,2,3,4, Jonathan D Olson1
1Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire.
Background:
In open-cranial neurosurgery, preoperative magnetic resonance (pMR) images are typically coregistered for intraoperative guidance. Their accuracy can be significantly degraded by intraoperative brain deformation, especially when resection is involved.
Objective:
To produce model updated MR (uMR) images to compensate for brain shift that occurred during resection, and evaluate the performance of the image-updating process in terms of accuracy and computational efficiency.
Methods:
In 14 resection cases, intraoperative stereovision image pairs were acquired after dural opening and during resection to generate displacement maps of the surgical field. These data were assimilated by a biomechanical model to create uMR volumes of the evolving surgical field. A tracked stylus provided independent measurements of feature locations to quantify target registration errors (TREs) in the original coregistered pMR and uMR as surgery progressed.
Results:
Updated MR TREs were 1.66 ± 0.27 and 1.92 ± 0.49 mm in the 14 cases after dural opening and after partial resection, respectively, compared to 8.48 ± 3.74 and 8.77 ± 4.61 mm for pMR, respectively. The overall computational time for generating uMRs after partial resection was less than 10 min.
Conclusion:
We have developed an image-updating system to compensate for brain deformation during resection using a computational model with data assimilation of displacements measured with intraoperative stereovision imaging that maintains TREs less than 2 mm on average.