Clinical experience with machine log file software for volumetric-modulated arc therapy techniques.
Luis Alberto Vazquez-Quino1, Claudia Ivette Huerta-Hernandez1, Dharanipathy Rangaraj1
1Radiation Oncology Department, Baylor Scott & White Health and Texas A&M Health Science Center College of Medicine, Temple, Texas.
Mobius FX software offers efficient quality assurance for intensity-modulated radiation therapy (IMRT) by verifying patient doses using treatment logs. This study found Mobius FX comparable to traditional methods like ArcCheck for VMAT plan QA.
Area of Science:
- Medical Physics
- Radiation Oncology
- Quality Assurance
Background:
- Intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) require rigorous quality assurance (QA) to ensure accurate patient dose delivery.
- Traditional QA methods can be time-consuming and may require specific device positioning.
- There is a need for efficient and accurate QA tools in radiation therapy.
Purpose of the Study:
- To evaluate the performance of Mobius FX, a novel software module for IMRT/VMAT QA.
- To compare the accuracy and efficiency of Mobius FX against traditional QA methods (ArcCheck) and the treatment planning system (TPS).
Main Methods:
- Thirty VMAT plans delivered on a Varian TrueBeam linac were analyzed.
- Mobius FX was used to calculate and verify 3D doses, utilizing linac treatment logs.
- Comparisons were made between Mobius FX, ArcCheck, and the TPS using gamma analysis (3%-3 mm and 2%-2 mm criteria).
Main Results:
- Mobius FX demonstrated high agreement with the TPS, with average gamma values of 99.96% (3%-3 mm) and 98.80% (2%-2 mm).
- Results from Mobius FX closely matched those obtained using traditional methods (ArcCheck).
- Profile analysis confirmed comparable results between Mobius FX and traditional QA methods.
Conclusions:
- Mobius FX provides a reliable and efficient method for pre-treatment patient-specific QA in VMAT.
- The software's ability to use linac treatment logs without device positioning enhances its practicality.
- Mobius FX is a viable alternative to traditional QA methods, offering comparable accuracy.