Related Experiment Video
Updated: Feb 27, 2026

05:58
Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
4.6K
Reticulocyte count and extended reticulocyte parameters by Mindray BC-6800: Reference intervals and comparison with
M Buttarello1, A Rauli2, G Mezzapelle2
1Department of Laboratory Medicine, Azienda Ospedaliera-Università di Padova, Padova, Italy.
International Journal of Laboratory Hematology
|July 5, 2017
Summary
The Mindray BC-6800 analyzer demonstrates good analytic performance for reticulocyte counts and extended parameters. Its diagnostic efficiency, particularly for mean reticulocyte hemoglobin equivalent (RHE), shows high agreement with the Sysmex XE-5000.
Area of Science:
- Hematology
- Clinical Chemistry
- Laboratory Medicine
Background:
- Evaluation of the Mindray BC-6800 analyzer for reticulocyte quantification.
- Assessment of extended reticulocyte parameters and their diagnostic utility.
- Determination of reference intervals in an adult population.
Purpose of the Study:
- To evaluate the analytic performance of the Mindray BC-6800 analyzer.
- To assess the diagnostic efficiency of reticulocyte parameters.
- To compare results with the Sysmex XE-5000 analyzer.
Main Methods:
- Analysis of imprecision, carryover, and time stability of the BC-6800.
- Inclusion of 184 healthy adults and 368 patients with various hematologic conditions.
- Comparison of reticulocyte and extended parameter measurements with the Sysmex XE-5000.
Main Results:
- Established reference intervals for reticulocytes, immature reticulocyte fraction, mean reticulocyte hemoglobin equivalent (RHE), and mean reticulocyte volume.
- Demonstrated low imprecision, negligible carryover (2.3%), and excellent time stability (up to 8 hours).
- Achieved good overall correlation for reticulocyte counting and 91.6% total agreement for RHE diagnostic efficiency.
Conclusions:
- The Mindray BC-6800 analyzer exhibits robust analytic performance for reticulocyte analysis.
- Good correlation exists between BC-6800 and XE-5000 for reticulocyte counts, with minor differences in upper reference limits.
- The BC-6800 shows high diagnostic efficiency for extended reticulocyte parameters, particularly RHE.