Repeatability of apparent diffusion coefficient and intravoxel incoherent motion parameters at 3.0 Tesla in orbital lesions
Related Experiment Video
Updated: Feb 27, 2026

Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring
Published on: December 9, 2010
Augustin Lecler1,2, Julien Savatovsky3, Daniel Balvay4
1Department of Radiology, Fondation Ophtalmologique Adolphe de Rothschild, 29 rue Manin, 75019, Paris, France. alecler@for.paris.
Objectives:
To evaluate repeatability of intravoxel incoherent motion (IVIM) diffusion-weighted imaging (DWI) parameters in the orbit.
Methods:
From December 2015 to March 2016, 22 patients were scanned twice using an IVIM sequence with 15b values (0-2,000 s/mm2) at 3.0T. Two readers independently delineated regions of interest in an orbital mass and in different intra-orbital and extra-orbital structures. Short-term test-retest repeatability and inter-observer agreement were assessed using the intra-class correlation coefficient (ICC), the coefficient of variation (CV) and Bland-Altman limits of agreements (BA-LA).
Results:
Test-retest repeatability of IVIM parameters in the orbital mass was satisfactory for ADC and D (mean CV 12% and 14%, ICC 95% and 93%), poor for f and D*(means CV 43% and 110%, ICC 90% and 65%). Inter-observer repeatability agreement was almost perfect in the orbital mass for all the IVIM parameters (ICC = 95%, 93%, 94% and 90% for ADC, D, f and D*, respectively).
Conclusions:
IVIM appeared to be a robust tool to measure D in orbital lesions with good repeatability, but this approach showed a poor repeatability of f and D*.
Key Points:
• IVIM technique is feasible in the orbit. • IVIM has a good-acceptable repeatability of D (CV range 12-25 %). • IVIM interobserver repeatability agreement is excellent (ICC range 90-95 %). • f or D* provide higher test-retest and interobserver variabilities.