Published on: December 9, 2010
Repeatability of apparent diffusion coefficient and intravoxel incoherent motion parameters at 3.0 Tesla in orbital
Augustin Lecler1,2, Julien Savatovsky3, Daniel Balvay4
1Department of Radiology, Fondation Ophtalmologique Adolphe de Rothschild, 29 rue Manin, 75019, Paris, France. alecler@for.paris.
Objectives:
To evaluate repeatability of intravoxel incoherent motion (IVIM) diffusion-weighted imaging (DWI) parameters in the orbit.
Methods:
From December 2015 to March 2016, 22 patients were scanned twice using an IVIM sequence with 15b values (0-2,000 s/mm2) at 3.0T. Two readers independently delineated regions of interest in an orbital mass and in different intra-orbital and extra-orbital structures. Short-term test-retest repeatability and inter-observer agreement were assessed using the intra-class correlation coefficient (ICC), the coefficient of variation (CV) and Bland-Altman limits of agreements (BA-LA).
Results:
Test-retest repeatability of IVIM parameters in the orbital mass was satisfactory for ADC and D (mean CV 12% and 14%, ICC 95% and 93%), poor for f and D*(means CV 43% and 110%, ICC 90% and 65%). Inter-observer repeatability agreement was almost perfect in the orbital mass for all the IVIM parameters (ICC = 95%, 93%, 94% and 90% for ADC, D, f and D*, respectively).
Conclusions:
IVIM appeared to be a robust tool to measure D in orbital lesions with good repeatability, but this approach showed a poor repeatability of f and D*.
Key Points:
• IVIM technique is feasible in the orbit. • IVIM has a good-acceptable repeatability of D (CV range 12-25 %). • IVIM interobserver repeatability agreement is excellent (ICC range 90-95 %). • f or D* provide higher test-retest and interobserver variabilities.