Image quality, diagnostic accuracy, and potential for radiation dose reduction in thoracoabdominal CT, using Sinogram
Michael Scharf1, Stephanie Brendel1, Katja Melzer1
1Department of Radiology, Friedrich-Alexander-University Erlangen-Nuremberg, Erlangen, Germany.
Objective:
To step-wise evaluate image quality of sinogram-affirmed iterative reconstruction (SAFIRE) in reduced-dose (RD) thoracoabdominal computed tomography (CT) compared to full-dose (FD) and RD filtered back projection (FBP) in a longitudinal study.
Materials And Methods:
122 patients were included in this prospective study. 49 patients (14 men: mean age ± SD, 56±0.4 years; 35 women: 58±1.3 years) completed FD, RD1 (80%-dose) and RD2 (60%-dose) thoracoabdominal CT. Each CT dataset was reconstructed with FBP and SAFIRE. For quantitative image analysis image noise was measured in defined tissue regions. Qualitative image evaluation was performed according to the European Guidelines on Quality criteria for CT. Additionally artifacts, lesion conspicuity, and edge sharpness were assessed.
Results:
Compared to FD-FBP noise in soft tissue increased by 12% in RD1-FBP and 27% in RD2-FBP reconstructions, whereas SAFIRE lead to a decrease of 28% (RD1) and 17% (RD2), respectively (all p <0.001). Visually sharp reproduction, lesion conspicuity, edge sharpness of pathologic findings, and overall image quality did not differ statistically significant between FD-FBP and RD-SAFIRE datasets. Image quality decreased in RD1- and RD2-FBP compared to FD-FBP, reaching statistically significance in RD2 datasets (p <0.001). In RD1- and RD2-FBP (p <0.001) streak artifacts were noted.
Conclusion:
Using SAFIRE the reference mAs in thoracoabdominal CT can be reduced by at least 30% in clinical routine without loss of image quality or diagnostic information.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...