Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
Ultrasound field characterization using synthetic schlieren tomography
Aki Pulkkinen1, Jarkko J Leskinen1, Aimo Tiihonen1
1Department of Applied Physics, University of Eastern Finland, P.O. Box 1627, Kuopio, 70211, Finland.
Abstract:
Synthetic schlieren imaging, also known as background oriented schlieren imaging, is used to determine the acoustical field of a focused ultrasound transducer operating at 1.01 MHz frequency with peak pressure amplitude of 0.97 MPa. The measurement setup is composed of a commercial off-the-shelf digital single-lens reflex (DSLR) camera with an ordinary objective, a high power light-emitting diode driven in pulsating mode, water tank, ultrasound transducer, rotation stage, and driving electronics. Measurements are performed in tomographic fashion by rotating the ultrasound transducer within the water tank and photographing an imaged target behind the ultrasound field. The photographs are processed with a Horn-Schunck-type algorithm, commonly used in optical flow analysis, in order to determine the deflection of light rays as caused by ultrasound field induced acousto-optic effect. Inverse Radon transform is then used, with the deflection data, to obtain three-dimensional spatial distribution of the pressure field gradient, from which an approximation of the ultrasonic pressure field is computed. The pressure field obtained with synthetic schlieren tomography is then compared to hydrophone measurements mainly qualitatively.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies II: Ultrasonography