Subharmonic and Endoscopic Contrast Imaging of Pancreatic Masses: A Pilot Study
Flemming Forsberg1, Maria Stanczak1, Andrej Lyshchik1
1Departments of Radiology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.
Objectives:
To use subharmonic imaging (SHI) to depict the vascularity of pancreatic masses compared to contrast-enhanced endoscopic ultrasound (EUS) and pathologic results.
Methods:
Sixteen patients scheduled for biopsy of a pancreatic mass were enrolled in an Institutional Review Board-approved study. Pulse-inversion SHI (transmitting/receiving at 2.5/1.25 MHz) was performed on a LOGIQ 9 system (GE Healthcare, Milwaukee, WI) with a 4C transducer, whereas contrast harmonic EUS (transmitting/receiving at 4.7/9.4 MHz) was performed with a radial endoscope (GF-UTC180; Olympus Corporation, Tokyo, Japan) connected to a ProSound SSD α-10 scanner (Hitachi Aloka, Tokyo, Japan). Two injections of the contrast agent Definity (Lantheus Medical Imaging, North Billerica, MA) were administrated (0.3-0.4 and 0.6-0.8 mL for EUS and SHI, respectively). Contrast-to-tissue ratios (CTRs) in the mass and an adjacent vessel were calculated. Four physicians independently scored the images (benign to malignant) for diagnostic accuracy and inter-reader agreement.
Results:
One patient dropped out before imaging, leaving 11 adenocarcinomas, 1 gastrointestinal stromal tumor with pancreatic infiltration, and 3 benign masses. Marked subharmonic signals were obtained in all patients, with intratumoral blood flow clearly visualized with SHI. Significantly greater CTRs were obtained in the masses with SHI than with EUS (mean ± SD, 1.71 ± 1.63 versus 0.63 ± 0.89; P = .016). There were no differences in the CTR in the surrounding vessels or when grouped by pathologic results (P > .60). The accuracies for contrast EUS and SHI were low (<53%), albeit with a greater κ value for SHI (0.34) than for EUS (0.13).
Conclusions:
Diagnostic accuracy of contrast EUS and transabdominal SHI for assessment of pancreatic masses was quite low in this pilot study. However, SHI had improved tumoral CTRs relative to contrast EUS.
More Related Videos
06:57Utilizing High Resolution Ultrasound to Monitor Tumor Onset and Growth in Genetically Engineered Pancreatic Cancer Models
Published on: April 7, 2018
14:56Application of Indocyanine Green Fluorescence Imaging Technology in Laparoscopic Duodenum-Preserving Pancreatic Head Resection
Published on: November 21, 2025
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Endoscopic Procedures V: ERCP
Patient...
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...