Receiver calibration and the nonlinearity parameter measurement of thick solid samples with diffraction and
Hyunjo Jeong1, Daniel Barnard2, Sungjong Cho1
1Department of Mechanical Engineering, Wonkwang University, Iksan, Jeonbuk 54538, Republic of Korea.
Ultrasonics
|July 11, 2017
Summary
This study introduces a precise method for measuring the nonlinearity parameter (β) in thick solid materials. Accurate corrections for diffraction and attenuation ensure reliable results, showing less than 3.2% variation across different sample thicknesses.
Area of Science:
- Materials Science
- Acoustics
- Nonlinear Ultrasonics
Background:
- Accurate measurement of the nonlinearity parameter (β) is crucial for material characterization.
- Existing methods often require complex receiver calibration and corrections for diffraction and attenuation.
Purpose of the Study:
- To develop and validate an accurate analytical and experimental technique for determining the nonlinearity parameter (β) in thick solid samples.
- To address challenges in receiver calibration, beam diffraction, and material absorption effects.
Main Methods:
- A simplified self-reciprocity calibration procedure for broadband receivers was developed.
- Analytical derivations for diffraction and attenuation corrections for fundamental and second harmonics were performed.
- Finite amplitude, through-transmission method was employed using aluminum alloy samples of varying thicknesses (4-12 cm).
Main Results:
- The proposed technique successfully determined the nonlinearity parameter (β) in thick solid samples.
- Systematic investigation revealed the significant impact of diffraction and attenuation corrections.
- With proper corrections, the variation in β across different sample thicknesses was reduced to less than 3.2%.
Conclusions:
- The presented technique provides accurate nonlinearity parameter (β) measurements in thick solids.
- Accounting for diffraction and attenuation is essential for reliable ultrasonic material characterization.
- The method demonstrates robustness and consistency across various sample thicknesses.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
1.1K
When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
1.1K
Instrument Calibration
904
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
904