Case report of unexplained hypocalcaemia in a slightly haemolysed sample
1Worcestershire Acute Hospitals NHS Trust, Worcester, United Kingdom.
Biochemia Medica
|July 12, 2017
Summary
A laboratory error involving sample contamination and hemolysis caused a false low calcium result in a healthy patient. This highlights the importance of recognizing pre-analytical issues in diagnostic testing.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Diagnostic Errors
Background:
- Routine blood tests are common for patients presenting with non-specific symptoms.
- Pre-analytical errors can significantly impact laboratory test results.
- Hemolysis is a frequent complication during sample collection and processing.
Observation:
- A young, healthy adult patient presented with flu-like symptoms.
- Initial blood analysis showed a falsely low calcium level due to sample hemolysis and contamination.
- Urgent repeat investigations revealed normal calcium levels.
Findings:
- The initial abnormal calcium result was a pre-analytical error, not a true clinical finding.
- Sample contamination, compounded by hemolysis, masked the true nature of the error.
- The laboratory's standard flagging system failed to identify the contamination due to the hemolysis complication.
Implications:
- This case underscores the critical need for robust pre-analytical quality control in clinical laboratories.
- Healthcare providers should be aware of potential interferences like hemolysis and contamination.
- Improved protocols are necessary to detect and prevent missed diagnostic errors stemming from sample quality issues.