Related Experiment Video
Updated: Feb 26, 2026

A Laser-induced Mouse Model of Chronic Ocular Hypertension to Characterize Visual Defects
Published on: August 14, 2013
Evaluation of three hand-held tonometers in normal canine eyes
K L Tofflemire1, C Wang1, J K Jens2
1Iowa State University, Department of Veterinary Clinical Sciences, College of Veterinary Medicine, 1600 South 16th Street, Ames, IA, 50011, USA.
Abstract:
Tonometry, an indirect measurement of intraocular pressure (IOP), is important for the diagnosis and management of glaucoma and uveitis. The aim of this study was to compare the performance of three hand-held tonometers in normal canine eyes. Eyes from cadavers of dogs without observable ocular disease were used to compare tonometric measurements with direct manometry over a range of 7.4-65mmHg. In vivo measurements using the three tonometers in both eyes of 12 healthy dogs were compared. All tonometers significantly underestimated manometric values both ex vivo and in vivo. One tonometer showed a small fixed bias over the range of IOP, whilst the other two tonometers had a negative proportional bias. The results of this study show that differences exist between handheld tonometers across the clinically relevant range of IOP, and that all underestimate manometric measurements.