Impact of Different Independent Dose Verification Software Programs for Secondary Check
Takeshi Yamazaki1, Masanobu Itano1, Satoru Ishibashi2
1Department of Radiology, Inagi Municipal Hospital.
Purpose:
A multi-institutional study was performed to identify the impact of different independent dose verification programs on independent dose verification software program.
Methods:
Data for 1,543 treatment fields were collected in three institutions. RADCALC and Simple MU Analysis (Simple MU) using the Clarkson-based algorithm were used. RADCALC needs the input of radiological path length (RPL) from radiotherapy treatment planning systems (RTPSs) (Eclipse or Pinnacle3). The Simple MU computes the RPL using CT images independently from the RTPSs. Ion-chamber measurements were performed for commissioning the two programs and the RTPSs. Next, the results of the two programs were compared to the RTPSs obtained in the clinically-approved plans in all three institutions.
Results:
The commissioning results showed ±1.5% variation in the ion-chamber measurements and there was slight difference between the institutions. The RADCALC (0.9±2.2%) and the Simple MU (1.7±2.1%) results showed a slight systematic difference. Pinnacle3 computed longer RPLs because it used CT-physical density tables. Thus, there was an impact on the accuracy in the treatment plans involving bone and other high-density materials.
Conclusion:
Dose calculation algorithms in different dose verification programs provided similar results. However, care must be taken because different RPL calculation methods in the RTPSs may affect dose difference between different independent dose verification programs by 1%.
More Related Videos
10:23Author Spotlight: A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
Published on: June 23, 2023
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Data Validation
Key parameters for method validation include:
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...