Analysis of polychlorodibenzo-p-dioxins and polychlorodibenzofurans in stationary source emissions in GC-MS/MS using hydrogen as the carrier gas
Related Experiment Video
Updated: Feb 26, 2026

Hydrogen Production and Utilization in a Membrane Reactor
Published on: March 10, 2023
Paolo Benedetti1, Ettore Guerriero1, Silvia Mosca1
1Institute of Atmospheric Pollution Research, Italian National Research Council, Rome, Italy.
Abstract:
The European and American methods for the determination of polychlorodibenzo-p-dioxins and polychlorodibenzofurans in stationary source emissions require a high-resolution mass spectrometer detector. At the same time, all of the 17 toxic 2,3,7,8-chlorine-substituted isomers have to be separated by an appropriate chromatographic method. The resulting analysis has long runtimes and sometimes a double-column run is needed, which results in a huge waste of time, money and manual labor. To obtain a fast and cost-effective but still reliable analytical system, a single-column gas chromatography with hydrogen as carrier gas and tandem mass spectrometry detection is proposed. The use of hydrogen as the carrier gas is a good choice to shorten runtimes and improve the chromatographic resolution, while tandem mass spectrometry is a well-known alternative for this analysis. All the chromatographic and mass spectrometric parameters fill the requirements of the reference methods with a 35% reduction of runtimes. The accuracy is tested with three fly ash samples of a proficiency interlaboratory test. A good correlation between the results is obtained (R2 = 0.992, slope = 0.9675), and no coelutions are noted. The system robustness is tested during 5 years of constant use and the maximal relative standard deviation of the relative response factor is 18.8%.
More Related Videos
07:49On-line Analysis of Nitrogen Containing Compounds in Complex Hydrocarbon Matrixes
Published on: August 5, 2016
11:44Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
Published on: March 6, 2016
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Types of Detectors-II
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...