Minimizing rubidium-82 tracer activity for relative PET myocardial perfusion imaging

Eline D Huizing1, Joris D van Dijk, Jorn A van Dalen

  • 1Departments of aNuclear Medicine, Isala HospitalbCardiologycMedical Physics, ZwolledMIRA: Institute for Biomedical Technology and Technical Medicine, University of Twente, Enschede, The Netherlands.

Abstract

Insights

Reducing rubidium-82 (Rb-82) activity for myocardial perfusion imaging (MPI) is feasible. An activity of 682 MBq provides reliable diagnostic outcomes and image quality, supporting clinical adoption.

Area of Science:

  • Nuclear Medicine
  • Cardiology
  • Radiopharmacy

Background:

  • Current rubidium-82 (Rb-82) activities for PET myocardial perfusion imaging (MPI) may exceed requirements.
  • Optimizing administered activity is crucial for patient safety and resource utilization.

Purpose of the Study:

  • To determine the minimum effective Rb-82 activity for reliable relative PET MPI.
  • To assess diagnostic accuracy and image quality at reduced Rb-82 activities.

Main Methods:

  • Analysis of 140 rest-stress MPI-PET scans from 28 patients.
  • Simulation of Rb-82 activities (852, 682, 511, 341 MBq) from list-mode data, compared to a 1023 MBq reference.
  • Evaluation of summed rest score, total perfusion deficit, image quality, and diagnostic interpretation.

Main Results:

  • Differences in summed rest score (≥3) occurred with 682, 511, and 341 MBq in 7%, 14%, and 18% of patients, respectively.
  • Significant image quality deterioration was observed only at 341 MBq (P<0.001).
  • Diagnostic interpretation of stress-rest scans was consistent between 682 MBq and 1023 MBq.

Conclusions:

  • A significant reduction in administered Rb-82 activity for relative MPI is achievable.
  • 682 MBq of Rb-82 offers reliable diagnostic outcomes and image quality.
  • This reduced activity level is suitable for clinical adoption in PET MPI.