Investigating the Robustness Neighborhood Gray Tone Difference Matrix and Gray Level Co-occurrence Matrix Radiomic
Usman Mahmood1, Aditya P Apte, Joseph O Deasy
1From the Departments of *Medical Physics and †Radiology, Memorial Sloan Kettering Cancer Center, New York, NY.
Objective:
The aim of this study was to determine if optimized imaging protocols across multiple computed tomography (CT) vendors could result in reproducible radiomic features calculated from an anthropomorphic phantom.
Methods:
Materials with varying degrees of heterogeneity were placed throughout the lungs of the phantom. Twenty scans of the phantom were acquired on 3 CT manufacturers with chest CT protocols that had optimized protocol parameters. Scans were reconstructed using vendor-specific standards and lung kernels. The concordance correlation coefficient (CCC) was used to calculate reproducibility between features. For features with high CCC values, Bland-Altman analysis was also used to quantify agreement.
Results:
The mean Hounsfield unit (HU) was 32.93 HU (141.7 to -26.5 HU) for the rubber insert and 347.2 HU (-320.9 to -347.7 HU) for the wood insert. Low CCC values of less than 0.9 were calculated for all features across all scans.
Conclusions:
Radiomic features that are derived from the spatial distribution of voxel intensities should be particularly scrutinized for reproducibility in a multivendor environment.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT