Novel Rotational Speed Modulation System Used With Venoarterial Extracorporeal Membrane Oxygenation
Noritsugu Naito1, Takashi Nishimura2, Kei Iizuka3
1Department of Cardiothoracic Surgery, Graduate School of Medicine, University of Tokyo, Tokyo, Japan.
Background:
Femoral venoarterial extracorporeal membrane oxygenation (VA-ECMO) is widely used to maintain blood flow in patients with cardiogenic shock. However, retrograde blood flow increases left ventricular (LV) afterload during femoral VA-ECMO. Additional support by means of an intraaortic balloon pump (IABP) alleviates LV afterload but is associated with significant adverse events. We previously developed a system for rotational speed modulation in synchrony with the native cardiac cycle, for use with implantable continuous-flow LV assist devices. Here, we aimed to evaluate whether our novel rotation speed modulation system can improve coronary artery flow and reduce LV during femoral VA-ECMO.
Methods:
VA-ECMO was installed by means of right atrial drainage and distal abdominal aortic perfusion in six adult goats. Cardiogenic shock was induced with β-adrenergic antagonist infusion. An IABP was placed in the descending aorta. LV stroke work, LV end-systolic pressure, and coronary arterial flow were evaluated. Data were collected under five conditions (modes): baseline, circuit-clamp (cardiogenic shock), continuous mode (constant rotational speed), counterpulse mode (increasing rotational speed during diastole), and continuous mode with IABP support.
Results:
LV stroke work and LV end-systolic pressure tended to be lower in the counterpulse mode, indicating decreased LV work load and afterload in this mode. Furthermore, coronary arterial flow tended to be higher in the counterpulse mode.
Conclusions:
Our system enabled an increase in coronary arterial flow and a decrease in LV work load and afterload during VA-ECMO. The system offers the effects of VA-ECMO and an IABP in a single device.
More Related Videos
03:40Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....