To retrieve, or not to retrieve: System revisions with the Micra transcatheter pacemaker
Eric Grubman1, Philippe Ritter2, Christopher R Ellis3
1Section of Cardiovascular Medicine, Yale University School of Medicine, New Haven, Connecticut.
Background:
Early experience with leadless pacemakers has shown a low rate of complications. However, little is known about system revision in patients with these devices.
Objective:
The purpose of this study was to describe the system revision experience with the Micra Transcatheter Pacing System (TPS).
Methods:
Patients with implants from the Pre-market Micra Transcatheter Pacing Study and the Micra Transcatheter Pacing System Continued Access Study (N = 989) were analyzed and compared with 2667 patients with transvenous pacemakers (TVPs). Revisions included TPS retrieval/explant, repositioning, replacement, or electrical deactivation (with or without prior attempt at retrieval, generally followed by TVP implant) for any reason. Kaplan-Meier revision rates were calculated to account for varying follow-up duration and were compared using a Fine-Gray competing risk model.
Results:
The actuarial rate for revision at 24 months postimplant was 1.4% for the TPS group (11 revisions in 10 patients), 75% (95% confidence interval 53%-87%; P < .001) lower than the 5.3% for the TVP group (123 revisions in 117 patients). TPS revisions occurred 5-430 days postimplant for elevated pacing thresholds, need for alternate therapy, pacemaker syndrome, and prosthetic valve endocarditis; none were due to device dislodgment or device-related infection. TPS was disabled and left in situ in 7 cases, 3 were retrieved percutaneously (range 9-406 days postimplant), and 1 was surgically removed during aortic valve surgery.
Conclusion:
The overall system revision rate for patients with TPS at 24 months was 1.4%, 75% lower than that for patients with TVPs. TPS was disabled and left in situ in 64% of revisions, and percutaneous retrieval was successful as late as 14 months postimplant.