Accuracy of densitometry of two cone beam computed tomography equipment in comparison with computed tomography
Sina Haghanifar1, Sajad Yousefi2, Ehsan Moudi1
1DDS., Associate Professor, Department of Oral and Maxillofacial Radiology, Faculty of Dentistry, Babol University of Medical Sciences, Babol, Iran.
Background:
The use of oral implants has been growing, and cone beam computerized tomography (CBCT) has become the method of choice for oral and maxillofacial radiology.
Objective:
To assess the accuracy of bone densitometry in two different CBCT devices in comparison with MDCT (multi-detector CT).
Methods:
Different concentrations of urografin, including 2.5%, 5%, 7.5%, 10%, 12.5%, were prepared, and the Hounsfield unit of these solution was measured by two CBCT devices (SORDEX CRANEX 3D and NEWTOM 5G) and one spiral CT device (SOMATOM SENSATION). Difference of output Hounsfield units in each concentration was compared in three devices. Correlation of devices with increase of urografin dose also was evaluated. Statistical analyses of the data were performed using SPSS18 and Kruskal-Wallis and Mann-Whitney U tests, along with Spearman's correlation coefficient.
Results:
The range of gray density for NEWTOM 5G CBCT, SORDEX 3D CBCT, and SOMATOM CT imaging systems was from 781 to 2311, 427 to 1464, and 222 to 994, respectively. There was significant difference between devices in the Hounsfield unit in all urografin concentrations (p<0.001). Also there was a significant correlation between three devices with increasing the urografin dose (p<0.05; r>0.95).
Conclusion:
Our findings indicated a high correlation and linear relationship between different studied imaging systems. Although utilizing CBCT in the assessment of bone density is useful according to its lower emitted dose and less cost, clinicians should be aware of the issue that the voxel value in CBCT is not as perfect as CT.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT