Detection of rotavirus in clinical specimens using an immunosensor prototype based on the photon burst counting
Makoto Hasegawa1, Ernest Apondi Wandera2, Yuka Inoue1
1Graduate School of Bioscience, Nagahama Institute of Bioscience and Technology, 1266 Tamura, Nagahama-shi, Shiga 526-0829, Japan.
Abstract:
In this study, a sensitive fluorescence sensor was developed for the detection of small, fluorescence-labeled particles dispersed in a solution. The prototype system comprises of a laser confocal optical system and a mechanical sample stage to detect photon bursting of fluorescence-labeled small particles in sample volumes less than 5 μL within 3 minutes. To examine the feasibility of the prototype system as a diagnostic tool, assemblages of rotavirus and fluorescence-labeled antibody were analyzed. The detection sensitivity for rotavirus was 1 × 104 pfu/mL. Rotavirus in stool samples from patients with acute gastroenteritis was also detected. The advantages and disadvantages of this immunosensor with respect to ELISA and RT-PCR, the current gold standards for virus detection, are discussed.