T2 -prepared balanced steady-state free precession (bSSFP) for quantifying whole-blood oxygen saturation at 1.5T
Ana E Rodríguez-Soto1, Osheiza Abdulmalik2, Michael C Langham1
1Laboratory for Structural, Physiologic and Functional Imaging, Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
Purpose:
To establish a calibration equation to convert human blood T2 to the full range of oxygen saturation levels (HbO2 ) and physiologic hematocrit (Hct) values using a T2 -prepared balanced steady-state free precession sequence (T2 -SSFP) at 1.5T.
Methods:
Blood drawn from 10 healthy donors (29.1 ± 3.9 years old) was prepared into samples of varying HbO2 and Hct (n = 79), and imaged using T2 -SSFP sequence at 37°C and interrefocusing interval τ180 = 12 ms. The relationship between blood T2 , HbO2 , and Hct was established based on the model R2=R2,plasma+Hct (R2,RBC-R2,plasma)+k·Hct·(1-Hct)·(1-HbO2)2. Measured R2 and HbO2 levels were fit by the model yielding values of R2,plasma, R2,RBC, and k. T2 -SSFP and the established calibration equation were applied to extract HbO2 at the superior sagittal sinus (SSS) in vivo and were compared with susceptometry-based oximetry.
Results:
Constants derived from the fit were: k = 74.2 [s-1 ], R2,plasma = 1.5 [s-1 ], R2,RBC = 11.6 [s-1 ], the R2 of the fit was 0.95. Average HbO2 at the SSS in seven healthy volunteers was 65% ± 7% and 66% ± 7% via T2 - and susceptometry-based oximetry, respectively. Bland-Altman analysis indicated agreement between the two oximetric methods with no significant bias.
Conclusion:
The calibration constants presented here should ensure improved accuracy for whole-blood oximetry based on T2 -SSFP at 1.5T. Magn Reson Med 79:1893-1900, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
More Related Videos
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
08:23Author Spotlight: Advancing Lung Disease Research with Free-Breathing Hyperpolarized Xenon-129 MRI
Published on: November 10, 2023
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Magnetic Resonance Imaging
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...