Performing In Vivo and Ex Vivo Electrical Impedance Myography in Rodents
Published on: June 8, 2022
Recording characteristics of electrical impedance myography needle electrodes
H Kwon1, S B Rutkove1, B Sanchez1
1Department of Neurology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA 02215-5491, United States of America.
Objective:
Neurologists and physiatrists need improved tools for the evaluation of skeletal muscle condition. Here we evaluate needle electrical impedance myography (EIM), a new minimally invasive approach to determine muscle status that could ultimately become a bedside tool for the assessment of neuromuscular disorders.
Approach:
We design and study the recording characteristics of tetrapolar EIM needle electrodes combining theory and finite-element model simulations. We then use these results to build and pilot in vivo an EIM needle electrode in the rat gastrocnemius muscle ([Formula: see text]). The dielectric properties of muscle are reported (mean ± standard deviation).
Results:
The numerical simulations show that the contribution of subcutaneous fat and muscle tissues to needle EIM data is <3% and >97%, respectively, and the sensed volume is [Formula: see text] cm3. Apparent resistivity [Formula: see text] [Formula: see text] cm and relative permittivity [Formula: see text] (dimensionless) measured at 10 kHz are in good agreement with in vivo dielectric properties reported in the literature.
Significance:
The results presented show the feasibility of measuring muscle impedivity in vivo using a needle electrode from 10 kHz to 1 MHz. The development of needle EIM technology can open up a new field of study in electrodiagnostic medicine, with potential applications to both disease diagnosis and biomarker assessment of therapy.