Related Experiment Video
Updated: Feb 26, 2026

Cardiac Magnetic Resonance Imaging at 7 Tesla
Published on: January 6, 2019
A dual flip angle 3D bSSFP magnetization transfer-like method to differentiate between recent and old myocardial
Philippe Germain1,2, Soraya El Ghannudi1,3, Aissam Labani1
1Department of Radiology, University Hospital, Strasbourg, France.
Background:
Magnetic resonance imaging (MRI) tissue signal is modulated by magnetization transfer (MT) phenomena, intrinsically induced by balanced steady-state free precession (bSSFP) imaging.
Purpose:
To investigate the possible value of such a MT-like bSSFP approach in two clinical settings involving focal myocardial lesions highligthed by late gadolinium enhancement (LGE+): edema induced by recent myocardial infarction (MI) and fibrotic scar related to chronic infarction.
Materials And Methods:
Population: 48 LGE + patients were studied: 26 with recent MI, 22 with chronic MI. 20 LGE-normal subjects were considered the control group. Field strength/sequence: Navigator-based short axis 3D-bSSFP sequences with 20° and 90° excitation flip angles were acquired (1.5T).
Assessment:
Pixel-wise normalized MT Ratio (nMTR) parametric images were calculated according to: nMTR = 100*(S20 -S90 *k)/S20 , with S20 and S90 signal intensity in 20° and 90° flip angle images and k = Blood20 /Blood90 as a normalization ratio. Statistical tests: analysis of variance (ANOVA), receiver operating characteristic (ROC) analysis.
Results:
Overall normal myocardial nMTR was 50.2 ± 3.6%. In recent MI, nMTR values were significantly reduced in LGE + regions (-22.3 ± 9.9%, P < 0.0001). In cases of chronic infarct, nMTR was significantly increased in LGE + regions (14.2 ± 11.4%, P < 0.0001). Comparison between observed results and theoretical values obtained with the Freeman-Hill formula showed that most variations observed in MI are related to MT effects instead of relaxation effects.
Conclusion:
In contrast to LGE imaging, which may show a similar hyperenhancement in recent and old infarctions, nMTR imaging demonstrates an opposite pattern: decreased values for recent infarction and increased values for old infarction, thus allowing to discriminate between these two clinical conditions without gadolinium injection.
Level Of Evidence:
3 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2018;47:798-808.