Improved Metrological Traceability of Particle Size Values Measured with Line-Start Incremental Centrifugal Liquid
Vikram Kestens1, Victoria A Coleman2, Pieter-Jan De Temmerman3
1Directorate-General Joint Research Centre, European Commission , 2440 Geel, Belgium.
Abstract:
Line-start incremental centrifugal liquid sedimentation (disc-CLS) is a powerful method to determine particle size based on the principles of Stokes' law. Because several of the input quantities of the Stokes equation cannot be easily determined for this case of a rotating disc, the disc-CLS approach relies on calibrating the sedimentation time scale with reference particles. To use these calibrant particles for establishing metrological traceability, they must fulfill the same requirements as those imposed on a certified reference material, i.e., their certified Stokes diameter and density value must come with a realistic measurement uncertainty and with a traceability statement. As is the case for several other techniques, the calibrants do not always come with uncertainties for the assigned modal diameter and effective particle density. The lack of such information and the absence of a traceability statement make it difficult for the end-user to estimate the uncertainty of the measurement results and to compare them with results obtained by others. We present the results of a collaborative study that aimed at demonstrating the traceability of particle size results obtained with disc-CLS. For this purpose, the particle size and effective particle density of polyvinyl chloride calibrants were measured using different validated methods, and measurement uncertainties were estimated according to the Guide to the Expression of Uncertainty in Measurement. The results indicate that the modal Stokes diameter and effective particle density that are assigned to the calibrants are accurate within 5% and 3.5%, respectively, and that they can be used to establish traceability of particle size results obtained with disc-CLS. This conclusion has a great impact on the traceability statement of certified particle size reference materials, for which the traceability is limited to the size and density values of the calibrant particles.
More Related Videos
08:52Improving Reproducibility to Meet Minimal Information for Studies of Extracellular Vesicles 2018 Guidelines in Nanoparticle Tracking Analysis
Published on: November 17, 2021
09:01An Ultra-clean Multilayer Apparatus for Collecting Size Fractionated Marine Plankton and Suspended Particles
Published on: April 19, 2018
Related Concept Videos
Centrifugation
Precipitate Formation and Particle Size Control
The obtained precipitate should be either a pure substance of known composition or easily converted to one by a simple process, such as ignition or drying. In addition, the precipitate should be insoluble and easily filterable. In general, filterability...