Handheld fetal Doppler device for assessing heart rate in neonatal resuscitation
Rinshu Shimabukuro1,2, Kozo Takase2, Sachiko Ohde3
1Department of Pediatrics, St Luke's International Hospital, Tokyo, Japan.
Background:
The use of neonatal three-lead electrocardiograms (Neo-ECG) is suggested for the assessment of heart rate (HR) in neonatal resuscitation, but primary childbirth facilities (i.e. maternity clinics or maternity homes) are not sufficiently equipped for this. The aim of this study was to therefore to confirm the potential of a handheld fetal Doppler device (FDD), widely used to assess HR in labor room resuscitation, and to compare its equivalency to Neo-ECG.
Methods:
In low-risk full-term infants, HR was measured five times per minute in minutes 1-5 after delivery using the FDD and Neo-ECG, and their equivalence compared. The equivalence margin was set at a HR difference of <±5 beats/min and <±5%. Moreover, the times at which HR was displayed were measured for both methods.
Results:
FDD and Neo-ECG HR correlated strongly (r = 0.977) in 155 measurements in 33 cases. The mean HR difference was -0.123 beats/min (95%CI: -0.594 to +0.345), and the logarithmic transformation of the HR ratio was -0.0003 (95%CI: -0.0020 to +0.0015). The mean HR difference and HR ratio were within the equivalence margin. For HR assessment using the FDD, 155 of 165 measurements (93.9%) could be obtained in <10 s.
Conclusions:
In HR assessment, FDD is a valid substitute for Neo-ECG in primary childbirth facilities that are not sufficiently equipped for Neo-ECG.
More Related Videos
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: