Related Experiment Video
Updated: Feb 26, 2026

A Murine Model of Stent Implantation in the Carotid Artery for the Study of Restenosis
Published on: May 14, 2013
Stent thrombosis with bioabsorbable polymer drug-eluting stents: insights from the Food and Drug Administration
Abdur R Khan1, Avnish Tripathi, Talha A Farid
1aDivision of Cardiovascular Medicine bDepartment of Internal Medicine, University of Louisville, Louisville, Kentucky cBrigham and Women's Hospital Heart and Vascular Center, Harvard Medical School, Boston, Massachusetts dDivision of Cardiology, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
Background:
SYNERGY, a bioabsorbable polymer-based, everolimus-eluting stent (BP-DES), recently received regulatory approval in the USA for use in percutaneous coronary interventions. Yet, information on the safety of BP-DES in routine clinical practice is limited. Our aim was to compare the safety of the recently approved BP-DES with current durable polymer drug-eluting stents (DP-DES) by analyzing adverse events, namely, stent thrombosis (ST), reported to the Manufacturer and User Facility Device Experience (MAUDE) database.
Materials And Methods:
The MAUDE database requires nationwide mandatory notification for adverse events on devices approved for clinical use. This database was searched for adverse events reported between 1 October 2015 and 25 December 2016, encountered after the placement of either BP-DES or DP-DES. Only those adverse events were included where the exposure period to the stents was comparable after the index procedure. Of all the adverse events reported, the event of interest was ST.
Results:
A total of 951 adverse events were reported. ST occurred in 48/951 of all events, 31/309 and 17/642 when BP-DES or DP-DES were used, respectively (P=0.00001). Of the 31 ST events with BP-DES, 68% (21/31) occurred within less than or equal to 24 h of the index procedure and 52% (16/31) occurred within less than or equal to 2 h.
Conclusion:
Our results raise the possibility of an increased risk of ST, particularly early ST (within 24 h), with the recently approved BP-DES. However, because of the inherent limitations of reporting within the MAUDE database, these data merely highlight a potential need for additional surveillance and randomized trials to assess further the safety of the bioabsorbable platform.
Related Concept Videos
Bioavailability Enhancement: Drug Stability Enhancement and GI Retention
Antiplatelet Drugs: Prostaglandin Synthesis, P2Y12 and Glycoprotein IIb/IIIa Inhibitors
Prostaglandin synthesis inhibitors, exemplified by the widely known aspirin, wield their power by irreversibly acetylating...
Anticoagulant Drugs: Vitamin K Antagonists and Direct Oral Anticoagulants
Warfarin, a prominent vitamin K antagonist family member, exerts its effect by inhibiting the enzyme VKORC1 (vitamin K epoxide reductase complex 1). By hindering this enzyme, warfarin...
Pharmacovigilance
This process, termed pharmacovigilance, aims to detect, evaluate, and minimize harmful effects related to medication use. The data collection for pharmacovigilance depends on spontaneous reporting systems, where healthcare professionals or patients voluntarily report suspected ADRs.
In some cases, there...
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Site-Targeted Drug Delivery Systems: Polymeric Carriers