Magnetic resonance imaging in patients with cardiac implantable electronic devices: a single-center prospective study
Mrinal Yadava1, Matthew Nugent1, Angela Krebsbach1
1Knight Cardiovascular Institute, Oregon Health and Science University, 3181 SW Sam Jackson Park Road, Portland, OR, 97239, USA.
Purpose:
Cardiac implantable electronic devices (CIEDs) have traditionally been a contraindication for magnetic resonance imaging (MRI). Recent studies suggest that MRI can be conducted safely in select patients with pacemakers (PPMs) and implantable cardioverter defibrillators (ICDs). We sought to determine the safety of MRI in patients with CIEDs, using a protocol for patient selection and device programming.
Methods:
This is a prospective, single-center study. Patients with a PPM or ICD and a clinical indication for MRI were considered. Exclusion criteria included newly implanted devices (<4 weeks), PPMs manufactured before 1996 and ICDs before 2000, epicardial and abandoned leads, and pacemaker-dependent ICD patients. Pacemaker-dependent PPM patients were programmed to asynchronous pacing. Tachycardia detection/therapies were disabled for ICDs. Devices were interrogated pre- and post-scan, and at follow-up 1-6 weeks later. Defibrillation threshold (DFT) was not tested post-scan. Patients were followed to monitor device therapies.
Results:
Two hundred twenty-seven patients underwent 293 scans. Devices included 170 (70.6%) PPMs and 71 (29.5%) ICDs. Thirteen (4.4%) scans were aborted mainly due to subjective complaints or artifact on scout cardiac imaging. Post-scan and follow-up interrogation demonstrated no changes in device parameters requiring reprogramming or revision. Over long-term follow-up (median, 354 days [IQR 65-629]), nine ICD patients had appropriate shocks (median, 3 [IQR 1-8]). One had four inappropriate shocks for atrial fibrillation. All tachyarrhythmias meeting criteria for defibrillation were successfully terminated.
Conclusions:
MRI can be conducted safely in patients with CIEDs when done in a protocoled manner with appropriate supervision. DFT testing after MRI may not be necessary.
More Related Videos
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT