Feasibility of optimized ultralow-dose pulsed fluoroscopy for upper gastrointestinal tract examinations: a phantom
Jakob Weiss1,2, Andreas Pomschar3, Carsten Rist3
1Department of Diagnostic and Interventional Radiology, Eberhard-Karls-University, University of Tuebingen, Hoppe-Seyler-Straße 3, 72076, Tuebingen, Germany. Jakob.Weiss@med.uni-tuebingen.de.
Purpose:
To establish an optimized ultralow-dose digital pulsed fluoroscopy (FP) protocol for upper gastrointestinal tract examinations and to investigate the radiation dose and image quality.
Materials And Methods:
An Alderson-Rando-Phantom with 60 thermoluminescent dosimeters was used for dose measurements to systematically evaluate the dose-area product (DAP) and organ doses of the optimized FP protocol with the following acquisition parameters: 86.7 kV; 77 mA; 0.9 mm3, automatic image noise and contrast adaption. Subjective image quality, depiction of contrast agent and image noise (5-point Likert scale; 5 = excellent) were assessed in 41 patients, who underwent contrast-enhanced FP with the aforementioned optimized protocol by two radiologists in consensus. A conventional digital radiograph (DR) acquisition protocol served as the reference standard for radiation dose and image quality analyses.
Results:
Phantom measurements revealed a general dose reduction of approximately 96% per image for the FP protocol as compared to the DR standard. DAP could be reduced by 97%. Significant dose reductions were also found for organ doses, both in the direct and scattered radiation beam with negligible orbital (FP 5.6 × 10-3 vs. DR 0.11; p = 0.02) and gonadal dose exposure (female FP 2.4 × 10-3 vs. DR 0.05; male FP 8 × 10-4 vs. DR 0.03; p ≤ 0.0004). FP provided diagnostic image quality in all patients, although reading scores were significantly lower for all evaluated parameters as compared to the DR standard (p < 0.05).
Conclusion:
Ultralow-dose FP is feasible for clinical routine allowing a significant reduction of direct and scattered dose exposure while providing sufficient diagnostic image quality for reliable diagnosis.
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Upper GI Series: Barium Swallow
Purpose and Procedure
Patients undergoing this procedure ingest a liquid containing barium sulfate with a chalky...
Imaging Studies III: Computed Tomography
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Imaging Studies II: Ultrasonography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...