Curtain Flow Column: Optimization of Efficiency and Sensitivity
Published on: June 12, 2016
Impact of instrument and column parameters on high-throughput liquid chromatography performance
Martin Gilar1, Thomas S McDonald1, Fabrice Gritti1
1Waters Corporation, 34 Maple Street, Milford, MA 01757, USA.
Abstract:
The speed and separation performance of high-throughput liquid chromatography (HT LC) was investigated. We evaluated the contributions of various experimental parameters to the total analysis cycle time, including column length (column void time), gradient delay, flow rate, auto-sampler (A/S) speed, and software related delays. The best case injection-to-injection cycle time of 22s was achieved using 12s gradient time and 2.1×20mm columns packed with 1.7μm C18 particles. The total 22s analytical duty cycle consisted of 2.5s column void time, 1.8s gradient delay, 12s gradient time, and approximately 5.7s for software setup delay time that served as column re-equilibration time. The achieved peak capacity for an alkylphenone sample was approximately 35, giving a peak production rate of 95 peaks per minute. We estimated the impact of LC system dispersion on peak capacity and peak production rate (peak capacity per unit of time). For HT LC scenarios (5-50mm columns and 4.8-120s long gradients) we observed that even a minor amount of system dispersion (2μL2) reduces the achievable peak capacity and peak productivity rate significantly. HT LC-MS analysis using 2.1×5mm guard column with duty cycle of 22s was successfully demonstrated.
Related Concept Videos
High-Performance Liquid Chromatography: Instrumentation
Principles Of Column Chromatography
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...