A New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device: Accuracy and User
Mark Christiansen1, Carmine Greene2, Scott Pardo2
11 Diablo Clinical Research, Walnut Creek, CA, USA.
Journal of Diabetes Science and Technology
|July 27, 2017
Summary
The Contour®Next ONE blood glucose monitoring system (BGMS) accurately syncs with a diabetes app. This system exceeded accuracy standards in both lab and clinical tests, demonstrating reliable glucose monitoring.
Area of Science:
- Medical Devices
- Diabetes Technology
- Clinical Diagnostics
Background:
- The Contour®Next ONE blood glucose monitoring system (BGMS) is a novel device designed for seamless synchronization with the Contour™ Diabetes app.
- Accurate blood glucose monitoring is crucial for effective diabetes management.
Purpose of the Study:
- To evaluate the analytical and clinical accuracy of the Contour®Next ONE BGMS.
- To assess user feedback on the ease of use of the BGMS and its accompanying app.
Main Methods:
- Laboratory accuracy was assessed using fingertip capillary blood samples against ISO 15197:2013 Section 6.3 standards.
- Clinical accuracy was evaluated in subjects with and without diabetes according to ISO 15197:2013 Section 8 criteria, comparing self-tests, alternate site tests, and venipunctures to a reference.
- User feedback was gathered through a questionnaire regarding system usability.
Main Results:
- 100% of laboratory results met ISO 15197:2013 Section 6.3 accuracy criteria across three test strip lots.
- In clinical studies, over 97% of self-test results met ISO 15197:2013 Section 8 accuracy criteria, with many within ±10% of the reference value.
- The majority of participants found the Contour®Next ONE system easy to use.
Conclusions:
- The Contour®Next ONE BGMS demonstrates high accuracy, exceeding established ISO 15197:2013 criteria in both laboratory and clinical settings.
- The system's performance suggests it is a reliable tool for blood glucose monitoring in individuals with diabetes.
- User feedback supports the system's ease of use, indicating good patient acceptance.