Three-Dimensional Reconstruction of Orbital Fractures
Published on: May 16, 2025
The validity and reliability of computed tomography orbital volume measurements
Silviu C Diaconu1, David Dreizin1, Mehmet Uluer1
1Division of Plastic, Reconstructive, and Maxillofacial Surgery, R Adams Cowley Shock Trauma Center, University of Maryland School of Medicine, 110S Paca St, Baltimore, MD, 21201, USA.
Purpose:
Orbital volume calculations allow surgeons to design patient-specific implants to correct volume deficits. It is estimated that changes as small as 1 ml in orbital volume can lead to enophthalmos. Awareness of the limitations of orbital volume computed tomography (CT) measurements is critical to differentiate between true volume differences and measurement error. The aim of this study is to analyze the validity and reliability of CT orbital volume measurements.
Materials And Methods:
A total of 12 cadaver orbits were scanned using a standard CT maxillofacial protocol. Each orbit was dissected to isolate the extraocular muscles, fatty tissue, and globe. The empty bony orbital cavity was then filled with sculpting clay. The volumes of the muscle, fat, globe, and clay (i.e., bony orbital cavity) were then individually measured via water displacement. The CT-derived volumes, measured by manual segmentation, were compared to the direct measurements to determine validity.
Results And Conclusions:
The difference between CT orbital volume measurements and physically measured volumes is not negligible. Globe volumes have the highest agreement with 95% of differences between -0.5 and 0.5 ml, bony volumes are more likely to be overestimated with 95% of differences between -1.8 and 2.6 ml, whereas extraocular muscle volumes have poor validity and should be interpreted with caution.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...