Characterization of Soft Contact Lens Fitting Using Ultra-Long Scan Depth Optical Coherence Tomography
Lele Cui1,2, Ming Li1,2, Meixiao Shen1,2
1School of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, China.
Objectives:
To evaluate the centration and movement of soft contact lenses and to verify the repeatability of two repeated measurements of the lens centration and movement using ultra-long scan depth optical coherence tomography (UL-OCT).
Methods:
A 1-day Acuvue® Define™ lens was tested on both eyes of 10 subjects (5 males and 5 females; mean age, 31.6 years). The centration and blink-induced movement of the contact lens were measured using UL-OCT at 5 min and 30 min after insertion. The measurements were repeated once at each checkpoint.
Results:
Good repeatability was found in the lens centration and movement between the two repeated measurements at either checkpoint. The values of the lens movement were 0.457 ± 0.248 mm and 0.402 ± 0.229 mm at 5 min and decreased to 0.197 ± 0.065 mm and 0.211 ± 0.110 mm at 30 min after insertion for the right and left eyes, respectively (P < 0.05).
Conclusions:
The custom-built UL-OCT presented good repeatability of centration and movement in Define lenses at 5 min and 30 min after insertion. Most of the lenses were centered temporal and inferior to the cornea during the first 30 min wearing period. Compared with 5 min after insertion, the lens was centered better and exhibited less movement at 30 min.