A Noncontact Dibutyl Phthalate Sensor Based on a Wireless-Electrodeless QCM-D Modified with Nano-Structured Nickel
Daqi Chen1, Xiyang Sun2, Kaihuan Zhang3
1State Key Laboratory of Industrial Control Technology, Institute of Cyber Systems and Control, Zhejiang University, Hangzhou 310027, China. 11332013@zju.edu.cn.
Abstract:
Dibutyl phthalate (DBP) is a widely used plasticizer which has been found to be a reproductive and developmental toxicant and ubiquitously existing in the air. A highly sensitive method for DBP monitoring in the environment is urgently needed. A DBP sensor based on a homemade wireless-electrodeless quartz crystal microbalance with dissipation (QCM-D) coated with nano-structured nickel hydroxide is presented. With the noncontact configuration, the sensing system could work at a higher resonance frequency (the 3rd overtone) and the response of the system was even more stable compared with a conventional quartz crystal microbalance (QCM). The sensor achieved a sensitivity of 7.3 Hz/ppb to DBP in a concentration range of 0.4-40 ppb and an ultra-low detection limit of 0.4 ppb of DBP has also been achieved.