Accuracy of eight deformable image registration (DIR) methods for tomotherapy megavoltage computed tomography (MVCT)
Wannapha Nobnop1,2, Hudsaleark Neamin1, Imjai Chitapanarux2
1Department of Radiologic Technology, Faculty of Associated Medical Sciences, Chiang Mai University, Chiang Mai, Thailand.
Introduction:
The application of deformable image registration (DIR) to megavoltage computed tomography (MVCT) images benefits adaptive radiotherapy. This study aims to quantify the accuracy of DIR for MVCT images when using different deformation methods assessed in a cubic phantom and nasopharyngeal carcinoma (NPC) patients.
Methods:
In the control studies, the DIR accuracy in air-tissue and tissue-tissue interface areas was observed using twelve shapes of acrylic and tissue-equivalent material inserted in the phantom. In the clinical studies, the 1st and 20th fraction MVCT images of seven NPC patients were used to evaluate application of DIR. The eight DIR methods used in the DIRART software varied in (i) transformation framework (asymmetric or symmetric), (ii) DIR registration algorithm (Demons or Optical Flow) and (iii) mapping direction (forward or backward). The accuracy of the methods was compared using an intensity-based criterion (correlation coefficient, CC) and volume-based criterion (Dice's similarity coefficient, DSC).
Results:
The asymmetric transformation with Optical Flow showed the best performance for air-tissue interface areas, with a mean CC and DSC of 0.97 ± 0.03 and 0.79 ± 0.11 respectively. The symmetric transformation with Optical Flow showed good agreement for tissue-tissue interface areas with a CC of (0.99 ± 0.01) and DSC of (0.89 ± 0.03). The sequences of target domains were significantly different in tissue-tissue interface areas.
Conclusions:
The deformation method and interface area affected the accuracy of DIR. The validation techniques showed satisfactory volume matching of greater than 0.7 with DSC analysis. The methods can yield acceptable results for clinical applications.
More Related Videos
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography