Related Experiment Video
Updated: Feb 25, 2026

Selected Reaction Monitoring Mass Spectrometry for Absolute Protein Quantification
Published on: August 17, 2015
Accurate quantification of 5 German cockroach (GCr) allergens in complex extracts using multiple reaction monitoring
S T Mindaye1, J Spiric1, N A David1
1Laboratory of Immunobiochemistry, Division of Bacterial, Parasitic, and Allergenic Products, Office of Vaccines Research and Review, CBER/FDA, Silver Spring, MD, USA.
Background:
German cockroach (GCr) allergen extracts are complex and heterogeneous products, and methods to better assess their potency and composition are needed for adequate studies of their safety and efficacy.
Objective And Methods:
The objective of this study was to develop an assay based on liquid chromatography and multiple reaction monitoring mass spectrometry (LC-MRM MS) for rapid, accurate, and reproducible quantification of 5 allergens (Bla g 1, Bla g 2, Bla g 3, Bla g 4, and Bla g 5) in crude GCr allergen extracts.
Results:
We first established a comprehensive peptide library of allergens from various commercial extracts as well as recombinant allergens. Peptide mapping was performed using high-resolution MS, and the peptide library was then used to identify prototypic and quantotypic peptides to proceed with MRM method development. Assay development included a systematic optimization of digestion conditions (buffer, digestion time, and trypsin concentration), chromatographic separation, and MS parameters. Robustness and suitability were assessed following ICH (Q2 [R1]) guidelines. The method is precise (RSD < 10%), linear over a wide range (r > 0.99, 0.01-1384 fmol/μL), and sensitive (LLOD and LLOQ <1 fmol/μL). Having established the parameters for LC-MRM MS, we quantified allergens from various commercial GCr extracts and showed considerable variability that may impact clinical efficacy.
Conclusions And Clinical Relevance:
Our data demonstrate that the LC-MRM MS method is valuable for absolute quantification of allergens in GCr extracts and likely has broader applicability to other complex allergen extracts. Definitive quantification provides a new standard for labelling of allergen extracts, which will inform patient care, enable personalized therapy, and enhance the efficacy of immunotherapy for environmental and food allergies.