Related Experiment Video
Updated: Feb 25, 2026

Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Addition of Nasal Cannula Can Either Impair or Enhance Preoxygenation With a Bag Valve Mask: A Randomized Crossover
David McQuade1, Matthew R Miller2, Clare Hayes-Bradley3
1From the Wellington Hospital, C&CDHB, Wellington, New Zealand.
Background:
A critical safety component of emergency anesthesia is the avoidance of hypoxemia during the apneic phase of a rapid sequence intubation. Preoxygenation with a bag valve mask (BVM) or anesthetic circuit may be improved with supplemental oxygen by nasal cannula (NC) if there is a mask leak. In addition, NC is recommended for apneic oxygenation after induction and may be placed before preoxygenation. However, the optimum NC flow rate for preoxygenation or whether the presence of NC alone creates a mask leak remains unclear.
Methods:
We performed a randomized crossover study on healthy volunteers comparing BVM alone and BVM with NC flow rates of 0 (NC-0), 5 (NC-5), 10 (NC-10), and 15 (NC-15) liters per minute (lpm). Our primary outcome was end-tidal oxygen (ETO2) after 3-minute preoxygenation.
Results:
There was no difference in ETO2 between NC-15, NC-10, or BVM-only at 3 minutes. NC-0 and NC-5 recorded significantly lower ETO2 at all times compared with NC-15, NC-10, or BVM-only (least difference NC-5, -7% [95% confidence interval {CI}, -4% to -10%), NC-0, 16% [95% CI, 13%-19%]). There was a difference in ETO2 between NC-15 and BVM-only at 1 minute (7%; 95% CI, 5%-9%), but not at 2 or 3 minutes. There was no difference in ETO2 between NC-10 and NC-15.
Conclusions:
Our study found that NC at 0 and 5 lpm with a BVM is deleterious to preoxygenation and should be avoided. In addition, a lack of difference between NC-10 and BVM-only demonstrates that NC at flows of at least 10 lpm should not impair the preoxygenation process. While NC-15 may offer a benefit by reaching maximal ETO2 at 1 minute, this would need to be balanced against patient comfort.
More Related Videos
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Cardiopulmonary Resuscitation II: ACLS Airway Management
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...