Performances and Limitations of Several Ultrasound-Based Elastography Techniques: A Phantom Study
Stéphanie Franchi-Abella1, Caroline Elie2, Jean-Michel Correas3
1Pediatric Radiology Department-Bicêtre Hospital, Hôpitaux Universitaires Paris-Sud- Assistance Publique Hôpitaux de Paris-Université Paris Sud, Le Kremlin-Bicêtre, France.
Abstract:
The objective of this study is to assess strain and shear wave (SW) elastography performance in terms of accuracy by performing in vitro measurements on a calibrated elastography phantom. Acquisitions were done on a phantom containing 4 inclusions (12-74 kPa) embedded in a homogeneous background material (30 kPa). We performed qualitative assessment on elastograms, semiquantitative assessment with strain or elasticity ratios between each inclusion and the background and quantitative evaluation with SW acquisitions. Ratio and elasticity estimations were compared with expected values. Biases, relative errors and 95% confidence intervals (95% CI) were calculated. All techniques adequately classified inclusions as harder or softer than the background. For stiffness ratio estimation, SW methods were more precise than strain methods and had significantly higher percentages of correctly classified measurements (p = 0.008). Quantitative stiffness measurements were reproducible despite constant biases. SW elastography methods provide more reproducible estimations of tissue stiffness ratio than strain methods, as well as reproducible quantitative tissue stiffness despite constant biases.
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasonography
During an ultrasonography procedure, a handheld device called...