The accuracy of static pressure measurement with water-filled urodynamic systems
1Bristol Urological Institute, Southmead Hospital, Bristol, United Kingdom.
Aims:
To quantify the inaccuracy of pressure measurement by water-filled systems, with a view to recommending better practice.
Methods:
Papers examining quality of measurements were reviewed, and standards documents examined. Technical data from manufacturers were gathered and summarized. Measurements were made on two different urodynamic systems. Individual inaccuracies were combined using the root sum of squares method.
Results:
The total inaccuracy in measurement may reach up to 9.6 cmH2 O on pdet , but with good practice this may be mitigated to 1.3 cmH2 O.
Conclusions:
Water-filled systems, while being the recommended method for urodynamic measurements, are subject to a potential inaccuracy during poor use of up to approximately 10 cmH2 O on pdet . Good practice can reduce this inaccuracy to a tolerable level, that is, approximately 1 cmH2 O.
More Related Videos
Related Concept Videos
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Static, Stagnation, Dynamic and Total Pressure
Pressure Variation in a Fluid at Rest
When measuring pressure at two different levels within the fluid, the difference in...
Urodynamic Studies: Uroflowmetry
Fluid Pressure
According to Pascal's law, a fluid at rest will generate equal pressure in all directions. This pressure is measured as a force per unit area, and its magnitude depends on the fluid's specific...
Pascal's Law