High-resolution mapping and ablation of recurrent left lateral accessory pathway conduction
Francesco Solimene1, Vincenzo Schillaci1, Gergana Shopova1
1Clinica Montevergine, Mercogliano, AV, Italy.
Abstract:
Proper localization of the anatomical target during ablation of the accessory pathways (AP) and the ability to detect clear AP potentials on the ablation catheter are crucial for successful AP ablation. We report a case of recurring AP conduction that was finally eliminated using a novel ablation catheter equipped with high-resolution mini-electrodes. Smaller and closer electrodes result in high mapping resolution with less signal averaging and cancellation effects. Owing to improved sensitivity, the new catheter seems effective in detecting fragmented and high frequency signals, thus allowing more effective radiofrequency application and improving ablation success.