Related Experiment Video
Updated: Feb 25, 2026

Evaluation of Blood Lactate and Plasma Insulin During High-intensity Exercise by Antecubital Vein Catheterization
Published on: May 18, 2018
Test results for the evaluation of a glucometer for use under hyperbaric conditions: Technical report
1Senior Biomedical Engineer, Department of Intensive Care and Hyperbaric Medicine, The Alfred Hospital, Melbourne, Australia.
Abstract:
This study aimed to evaluate a recently developed equipment test method by assessing the safe and accurate functioning of the Abbott Optium FreeStyle H portable blood glucose monitor for use in the Alfred Hospital's hyperbaric chamber. The results of this study indicate that the test method can be used successfully to evaluate instruments and/or devices for use in the hyperbaric environment. The evaluation initially found that this particular glucose monitor contained a lithium battery which can be hazardous when used in the hyperbaric environment. However, upon further inspection it was determined the battery posed minimal risk for fire and explosion due to its small capacity and design application. The results indicate that the Abbott Optium FreeStyle H blood glucose monitor operated normally when used in the hyperbaric chamber. This glucometer was found to perform within the calibration specification requirements for accuracy at all stages of a typical hyperbaric treatment and as such the Abbott Optium FreeStyle H blood glucose monitor was deemed safe for use in the hyperbaric chamber at the Alfred Hospital.

